| dc.contributor.advisor | Jaa'frah, Fidaa | |
| dc.contributor.author | AbuRajab, Haneen | |
| dc.contributor.author | Shalaldeh, Sahar | |
| dc.contributor.author | Odeh, Sally | |
| dc.date.accessioned | 2021-09-15T08:14:27Z | |
| dc.date.accessioned | 2022-05-22T08:01:23Z | |
| dc.date.available | 2021-09-15T08:14:27Z | |
| dc.date.available | 2022-05-22T08:01:23Z | |
| dc.date.issued | 2021-06-01 | |
| dc.identifier.uri | http://localhost:8080/xmlui/handle/123456789/7607 | |
| dc.description | CD, no of pages 87 , اجهزة طبية 4/2021 | |
| dc.description.abstract | Defibrillator analyzer is one of the most important calibration devices that is specified to test quality and functionality of defibrillators, in order to avoid injuries that may occur to patient thus to a higher or a lower dose of DC shock that has been obtained initially from defibrillator device. This project aims to design a low cost defibrillator analyzer, that depends on the received joule which is delivered by defibrillator, and attenuating the voltage to a sensible ratio that will be conditioned and programmed as an energy value that will be compared to the received energy from defibrillator in order to test the quality of the defibrillator for any need for calibration or maintenance | en_US |
| dc.language.iso | en | en_US |
| dc.publisher | جامعة بوليتكنك فلسطين - اجهزة طبية | en_US |
| dc.subject | Defibrillator analyzer | en_US |
| dc.title | Design and Implementation of Defibrillator analyzer | en_US |
| dc.type | Other | en_US |