DSpace Repository

A real time ionic dialysance monitoring by using conductivity cell

Show simple item record

dc.contributor.advisor jaafrah, Fida'a
dc.contributor.author Zamareh, Alaa
dc.contributor.author Alnajjar, Musab
dc.date.accessioned 2020-06-08T09:49:59Z
dc.date.accessioned 2022-05-22T06:49:13Z
dc.date.available 2020-06-08T09:49:59Z
dc.date.available 2022-05-22T06:49:13Z
dc.date.issued 2014-05-01
dc.identifier.uri http://localhost:8080/xmlui/handle/123456789/7102
dc.description no of pages 65, 27311, اجهزة طبية 9/2014 , in the store
dc.description.abstract During dialysis treatments, there is an interest of measuring the dialysis efficiency for each treatment , an easy method has been developed and is based on measuring the outlet conductivity of the dialysis fluid. The conductivity measuring continuously along dialysis session every 30 minutes . The project made for Fresenius hemodialysis machine model ( model 4008B) , and 4 – poles conductivity cell is used in the project , because it has high accuracy, and AC constant current is applied to it. en_US
dc.language.iso en en_US
dc.publisher جامعة بوليتكنك فلسطين - اجهزة طبية en_US
dc.subject conductivity cell en_US
dc.subject ionic dialysance en_US
dc.title A real time ionic dialysance monitoring by using conductivity cell en_US
dc.type Other en_US


Files in this item

This item appears in the following Collection(s)

Show simple item record

Search DSpace


Browse

My Account